FD800 Dual Control Moisture Balance

Flagship Moisture Balance — Dual Temperature Control for Complex, Scorch-Prone Samples

The top of Kett’s loss-on-drying line, built for the materials that defeat ordinary balances:  liquids, pastes, powders, and large particulates. A second, non-contact sensor reads the sample surface directly, so heat-sensitive and volatile samples dry fast without burning. On the proven FD720 platform, including the large 120 g capacity, six drying modes, patented auto-taring, the FD800 handles a demanding mix of materials with primary-method accuracy.

  • Dual temperature control: A chamber thermistor plus a non-contact radiation thermometer that controls to the sample surface up to 250 °C.  This is aggressive heat for speed, eased down as the surface dries.
  • Predictive measuring mode: Compares projected against actual weight change and ends the test once results meet confidence limits resulting in maximum throughput without cutting a reading short.
  • Guaranteed accuracy from a primary method: Loss on drying measures mass loss directly; run to completion, the result is definitive and stands up to scrutiny.
  • Stored profiles and results memory: Ten stored measuring conditions for instant recall across materials, plus on-board memory for 100 results.

$5,495.00

Description

Direct Surface-Temperature Control for Samples That Scorch Before They Dry

Conventional moisture balances sense only the air temperature inside the drying chamber. Sample depth, particle size, and volatility mean the surface can scorch which drives off compounds other than water and corrupting the result. More importantly, this can happen long before the chamber reading or visual inspection suggests a problem. The FD800 adds a second, non-contact sensor that reads the sample surface directly, giving you control over the temperature that actually governs the dry.

Flagship Capabilities

  • Dual temperature control: Alongside the chamber thermistor (T1), a non-contact radiation thermometer (T2) reads the sample surface directly and controls to it up to 250 °C. Apply high initial heat for fast drying, then feather it down as the surface dries.  The net result is removing moisture without scorching the sample or driving off other volatiles. No other moisture balance in the line sees the surface temperature.
  • Predictive measuring mode: The FD800 compares projected weight change against the actual change in real time and ends the test the moment results fall within confidence limits.  This unique feature ensures maximum throughput with no risk of cutting a measurement short. It is one of six drying modes, so you can match the profile to any sample.
  • Ready for high-mix workloads:  Ten stored measuring conditions keep a tuned, ready-to-recall profile for every material you test, and on-board memory holds 100 results for batch review or output.
  • Documentation built for GMP and GLP: Each result outputs over RS-232C as tab-delimited data straight into Microsoft Excel or a thermal printer.  There is no proprietary software required.  Password protection restricts settings changes, supporting the data-integrity and access-control practices these frameworks require.

Built on the Proven FD720 Platform

Beneath the dual-sensor system is the full FD720 platform which includes the large 120 g capacity and 130 mm dish along with Kett’s patented auto-taring to hold readings drift-free through long runs. Because loss on drying is a primary gravimetric method, a result run to completion is definitive and traceable, not inferred from a correlation. The FD800 handles liquids, pastes, powders, and large particulates that defeat single-sensor balances, while operating as directly as the rest of the line: set your conditions, place the sample, close the lid, and press start. Every unit carries Kett’s One-Year Total Care Warranty covering parts, labor, and specification compliance, with Maintenance Agreements available.

Don’t need non-contact surface control? The FD720 Universal Moisture Balance covers the same 120 g, six-mode loss-on-drying platform for everyday QC at a lower cost.

Testing routine samples or tight on bench space? The FD660 Compact Moisture Balance delivers efficient loss-on-drying analysis in a smaller, lighter package.

Additional information

Weight 19 lbs
Dimensions 22 × 30 × 15 in
Measurement Principle

Loss on Drying (heat-drying / weight-loss method)

Measurement Range

0-100% (wet base, solids); 0-500% (dry base)

Resolution

Moisture 0.01% or 0.1% (selectable); weight 0.001 g

Reproducibility

0.05% for samples 5 g and over; 0.02% for samples 10 g and over

Response Time

30 seconds; total drying time sample dependent

Applications

Solids, slurries, liquids, and mixes

Measurable Components

Moisture (wet base and dry base), solid content, weight

Form Factor

Benchtop

Sample Capacity

0.5-120 g

Sensor Type

Thermistor (T1, chamber) plus non-contact radiation thermometer (T2, sample surface), fixed emissivity 0.95

Display

Backlit LCD, 137 × 43 mm

Measurement Modes

Six: Automatic; Timed (1-240 min, or continuous to 12 hours); High-speed; Low-speed; Stepped (up to 5 steps); Predictive

Features

Dual temperature control (T1 thermistor plus T2 non-contact radiation thermometer) dries scorch-prone samples without burning; UniBloc weight sensor; auto-taring; ten-key entry pad

Data Memory Capacity

100 measurements

Test Protocol Storage

100 stored measurement-condition sets

Heating Method

Infrared heat drying

Heating Temperature Range

T1 30-180 °C; T2 30-250 °C, in 1 °C steps

Heat Source

Mid-infrared quartz lamp (2.6 µm), 200 W × 2

Operating Temperature

5-40 °C

Operating Humidity

85% RH maximum

Communications

RS-232C (2400 bps, 8 data bits, no parity, 1 stop bit, D-SUB9)

Power

AC 100-120 V / 220-240 V (50/60 Hz)

Power Consumption

900 W maximum

Dimensions

220 × 415 × 220 mm

Weight

5.4 kg (11.9 lbs)

Standard Accessories

SUS sample dishes, 130 mm dia × 13 mm deep (2); dish handlers (2), wind shield, dish tray, spoon and spatula set, spare fuses (2), aluminum foil sheets (20), glass fiber sheets (10), operating manual

Options

VZ380 printer set (VZC69 cable) or VZ330 set (VZC14 cable); GF200 radiation thermometer calibration kit; FDL01 data logger software; TQ100 sample crusher; FW100 case; foil sheets (500)

Warranty

One Year Total Care Warranty covering parts, labor, and specification compliance. Maintenance Agreements available.