Welcome to Kett - Tomorrow's Testing Today

USD|ENG

FD720 Universal Moisture Balance

Loss-on-Drying Moisture Balance — Six Drying Modes, Patented Auto-Taring, 0.001 g Resolution

Determine moisture and solid content by loss on drying — the primary gravimetric reference method that official methods and public regulations are built on. The FD720 handles the full range of QC samples, from grains and foods to chemicals, polymers, and pastes, with six drying modes to match the method to the material and a patented auto-taring system that holds readings steady through long runs.

  • Six measurement modes: Automatic, timed, high-speed, low-speed, stepped, and predictive drying — choose the profile that dries each material completely without scorching.
  • Guaranteed accuracy from a primary method: Loss on drying measures mass loss directly. Run to completion, the result is definitive and stands up to scrutiny — not an estimate calibrated against something else.
  • Patented auto-taring: Kett’s internal auto-taring re-zeros the balance during the test, eliminating the heat-driven scale drift that affects ordinary balances.
  • 120 g capacity: Accommodates large or low-density samples that overflow the small pans on compact analyzers.

$3,195.00

Description

The Advanced Professional Standard for Precision Moisture Analysis

The FD720 brings laboratory loss-on-drying — the primary gravimetric reference method that official methods and public regulations are built on — to the production QC bench, with the premium drying control and balance stability of higher-end analyzers. It suits chemicals, pharmaceuticals, foods, pulp and paper products, polymers, and most materials containing organic substances.

Solving Critical Laboratory Pain Points

  • Eliminating heater maintenance: The mid-infrared quartz heating lamps average five to ten times the operating life of halogen or older infrared elements.  With a MTBF of roughly 20,000 to 30,000 hours of measured use, this feature lowers total cost of ownership by reducing element replacements.
  • Guaranteed accuracy through long tests: Heat buildup causes scale drift on ordinary balances. Kett’s patented internal auto-taring system re-zeros the balance automatically throughout a run, keeping readings drift-free even on continuous tests of several hours. Because loss on drying is a primary gravimetric method, a result run to completion is definitive and traceable and not a value inferred from a correlation.
  • Documentation built for GMP and GLP: Each result outputs over RS-232C as tab-delimited data straight into Microsoft Excel or a thermal printer.  There is no proprietary software required.  Password protection restricts settings changes, supporting the data-integrity and access-control practices these frameworks require.
  • One instrument for varied samples: Six drying modes: automatic, timed, high-speed, low-speed, stepped, and predictive plus 120 g capacity and a 130 mm sample dish handle scorch-prone, bulky, or low-density materials that overflow the small pans on compact analyzers.

Built for Everyday QC

Operation is direct: set time and temperature, place the sample, close the lid, and press start. Ten stored measuring conditions recall the right profile for each material, and on-board memory holds 100 results. Every unit carries Kett’s One Year Total Care Warranty covering parts, labor, and specification compliance, with Maintenance Agreements available for calibration and ongoing support.

Drying samples that scorch before they read dry? The FD800 Dual Control Moisture Balance adds a non-contact sensor that reads sample-surface temperature directly, drying heat-sensitive materials quickly without burning them.

Don’t need 120 g capacity or six drying modes? The FD660 Compact Moisture Balance covers everyday loss-on-drying testing in a smaller, lower-cost package.

Additional information

Weight 18.2 lbs
Dimensions 22.5 × 11.5 × 14.5 in
Measurement Principle

Loss on Drying (heat-drying / weight-loss method)

Measurement Range

0-100% (wet base, solids); 0-500% (dry base)

Resolution

Moisture 0.01% or 0.1% (selectable); weight 0.001 g

Reproducibility

0.05% for samples 5 g and over; 0.02% for samples 10 g and over

Response Time

30 seconds; total drying time sample dependent

Applications

Solids, slurries, liquids, and mixes

Measurable Components

Moisture (wet base and dry base), solid content, weight

Form Factor

Benchtop

Sample Capacity

0.5-120 g

Sensor Type

Thermistor (T1, chamber)

Display

Backlit LCD, 137 × 43 mm

Measurement Modes

Six: Automatic; Timed (1-240 min, or continuous to 12 hours); High-speed; Low-speed; Stepped (up to 5 steps); Predictive

Features

UniBloc single-unit aluminum weight sensor; auto-taring eliminates scale drift; six measurement modes; mid-infrared quartz lamp, 20,000-30,000 hours typical; 30-second moisture-change display

Data Memory Capacity

100 measurements

Test Protocol Storage

10 stored measurement-condition sets

Heating Method

Infrared heat drying

Heating Temperature Range

30-180 °C in 1 °C steps

Heat Source

Mid-infrared quartz lamp (2.6 µm), 200 W × 2

Operating Temperature

5-40 °C

Operating Humidity

85% RH maximum

Communications

RS-232C (2400 bps, 8 data bits, no parity, 1 stop bit, D-SUB9)

Power

AC 100-120 V / 220-240 V (50/60 Hz)

Power Consumption

900 W maximum

Dimensions

220 × 415 × 190 mm

Weight

4.5 kg (9.9 lbs)

Standard Accessories

SUS sample dishes, 130 mm dia × 13 mm deep (2); dish handlers (2), wind shield, dish tray, spoon and spatula set, spare fuses (2), aluminum foil sheets (20), power cord, operating manual

Options

VZ380 printer set (VZC69 cable) or VZ330 set (VZC14 cable); VZC52 RS-232C cable; KDL01 data logger software; GF200 temperature calibration kit; TQ100 sample crusher; FW100 case; foil sheets (500)

Warranty

One Year Total Care Warranty covering parts, labor, and specification compliance. Maintenance Agreements available.

Reviews

There are no reviews yet.

Be the first to review “FD720 Universal Moisture Balance”

Your email address will not be published. Required fields are marked *

The questions is empty.